Human Endoplasmic Reticulum Aminopeptidase 2 (ERAP2) ELISA Kit
Principle of the Assay
The microtiter plate provided in this kit has been pre-coated with an antibody specific to ERAP2. Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody preparation specific to ERAP2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After the TMB substrate solution is added, only those wells that contain ERAP2, biotin-conjugated antibody, and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution, and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of ERAP2 in the samples is then determined by comparing the O.D. of the samples to the standard curve.
For Use with serum, plasma, and cell culture supernatants. For Research Use Only. Not for use in diagnostic procedures.
Target Information
Aminopeptidase that plays a central role in peptide trimming, a step required for the generation of most HLA class I-binding peptides. Peptide trimming is essential to customize longer precursor peptides to fit them to the correct length required for presentation on MHC class I molecules. Preferentially hydrolyzes the basic residues Arg and Lys.
GENE ID | 64167 |
SWISS PROT | Q6P179 |
SYNONYMS |
L-RAP; LRAP; Leukocyte-Derived Arginine Aminopeptidase |
Materials Supplied
Kit Components | 96 Wells Quantity/Size |
---|---|
Pre-coated, ready-to-use 96-well strip plate | 1 plate |
Plate sealer for 96 wells | 2 |
Standard |
2 tubes |
Diluent buffer | 1 bottle |
Detection Reagent A | 1 bottle |
Detection Reagent B | 1 bottle |
TMB Substrate | 1 tube |
Stop Solution | 1 tube |
Wash Buffer (30 ℅ concentrate) | 1 tube |
Product data sheet | 1 copy |
Storage
Storage | The TMB Substrate, Wash Buffer (30X concentrate), and the Stop Solution should be stored at 4°C upon receipt, while the other items should be stored at -20°C. |
Performance Characteristics
REPEATABILITY |
Intra-assay Precision (Precision within an assay): 3 samples with low, middle, and high-level ERAP2 were tested 20 times on one plate, respectively. |
SENSITIVITY | The minimum detectable dose was 0.115ng/mL. |
ASSAY RANGE | 0.312-20ng/mL |
SPECIFICITY | This assay has high sensitivity and excellent specificity for the detection of ERAP2. No significant cross-reactivity or interference between ERAP2 and analogs was observed. Note: Limited by current skills and knowledge, it is impossible to perform all possible cross-reactivity detection tests between ERAP2 and all analogs, therefore, cross reactivity may still exist. |